- 전자선 차폐물질에 의한 선량분포의 변화
- ㆍ 저자명
- 추성실
- ㆍ 간행물명
- 대한방사선의학회지
- ㆍ 권/호정보
- 1979년|15권 1호|pp.278-282 (5 pages)
- ㆍ 발행정보
- 대한영상의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
We studied for the shielding materials of normal tissue in high energy electron irradiation. The lead plate being choosed as the most appropriate shielding material and the dose distribution is transformed into following : 1) The thickness of lead for complete shielding of high energy electron is 3.2 mmPb for 8 MeV 4.3 mmPb for 10 MeV 5.0 mmPb for 12 MeV 2) Surface dose is increased about 3 to 10 percent of maximum dosage, if the lead plate is used for electron beam shielding. 3) The bremsstrahlung is occured from shielding lead plate from above 4 to 6 percent of maximum dosage, yet this dosage is not above the practical error of radiotherapy.