- 2차 샘플링을 이용한 2-D 도플러 시스템의 시뮬레이션과 시스템구현에 관한 연구
- ㆍ 저자명
- 임춘성,임용곤
- ㆍ 간행물명
- 의공학회지
- ㆍ 권/호정보
- 1990년|11권 1호|pp.147-156 (10 pages)
- ㆍ 발행정보
- 대한의용생체공학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
A two-dimensional pulsed doppler system for ultrasonic blood velocity doppler signals is studied and implemented. The second-order sampling method and serial data processing procedures are utillized in the sys- tem, which eliminates the untuning problems at phase channels in the quadrature detection method as well as in the channels of parallel data processing. rho digital signal processor used in this system allows a hardware savings and flexible design options. The efficiency of the various mean frequency estimators in the second-order sampling system is examined by computer simulation as a function of the intersequence sample delay time. The temporal delay for the quadrature component is changed from $1/(4f_o){;}to{;}3/(4f_o){;}and{;}5/(4f_o)$ where to is the center frequency of the transducer, It is found that autocorrelator is the optimum frequency estimator for the second-order sampling: with !he intersequence sample delay of $1/(4f_o){;}to{;}3/(4f_o){;}and{;}5/(4f_o)$. The qualitative variation and information proportional to blood velocity in the vessel system are obtained in the VIVO experiments.