- 대향2문조사시 Target Volume의 위치에 따른 Beam Weight의 최적화
- ㆍ 저자명
- 이진국,김지한,임익수,최영헌,Lee. Jin Guk,Kim. Ji Han,Im. Ik Su,Choe. Yeong Heon
- ㆍ 간행물명
- 大韓放射線治療技術學會誌
- ㆍ 권/호정보
- 1992년|5권 1호|pp.68-73 (6 pages)
- ㆍ 발행정보
- 대한방사선치료학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
If the same weight is used in parallel opposed 6 and 10 MV x-ray beams, the lowest dose is achieved at SAD. Therefore, dose homogeneity in the target volume is decreased when SAD is taken at center of target volume than center of phantom or patient. With Standard deviation of ${pm}6\%$ that repuesented the dose homogeneity in tarhet volume, we studied the optimized beam weights at which hot spot dose was least in parallel opposed beams. The optimized beam weights that maximally decrease the hot spot dose, wer 1.29, 1.19, 2.71, 3.50, and 4.70 in 6 MV x-ray and 1.25, 1.53, 1.90, 2.36, 3.01, and 3.7 in 10 MV x-ray, reapectively, when center of target volume was changed to 2,4,6,8,10, and 12cm from center plan of phantom along the centeral axis of beams.