치과 임플랜트 시술시 골천공기구의 회전속도가 주위 골조직의 온도 및 골일체성에 미치는 영향에 관한 연구
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • 치과 임플랜트 시술시 골천공기구의 회전속도가 주위 골조직의 온도 및 골일체성에 미치는 영향에 관한 연구
저자명
이진걸,양재호,이선형,Lee. Jin-Geol,Yang. Jae-Ho,Lee. Sun-Hyung
간행물명
대한치과보철학회지
권/호정보
1992년|30권 2호|pp.167-189 (23 pages)
발행정보
대한치과보철학회
파일정보
정기간행물|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Frictional heat produced by cutting tools during dental implantation procedure may destroy the surrounding bone tissue and regenerative capacity and interfere ossointegration by formation of undifferentiated connective tissue. To study the effect of frictional heat according to various rotational speeds on the regenerative capacity of surrounding bone tissue, 13 ITI HS implants (8 mm) were inserted at 4 mongrel dogs. Temperatures were measured using thermocouple located 6 mm below from the marginal crest and 0.5 mm from the periphery of trephine mill during implant bed preparation. After 4 and 9 months, animals were sacrificed and specimens were examined using x-rays and light microscope. Results were as follows: 1. With drill speeds of 300, 800, 2,000, 3,500 rpm and saline irrigation, temperatures of surrounding bone were decreased from $-2.9^{circ};to;-1.7^{circ}C$. Temperature rises of $2.0^{circ};and;2.1^{circ}C$ were recorded with a drill speed of 5,000 rpm and irrigation. 2. With drill speeds of 800, 3,500, 5,000 rpm and no irrigation, temperatures of surrounding bone rose from $+1.5^{circ};to;+6.8^{circ}C$, but maximum temperature was $40^{circ}C$ at 5,000 rpm. 3. On radiographic examination, bone resorptions were observed at the upper half of implant of 5,000 rpm without irrigation and one case of 5,000 rpm with irrigation. 4. Osseointegration was unsuccessful in cases of 3,500, 5,000 rpm without irrigation due to fibrous connective tissue formation to the outer surface and hollow, but it was successful in a case of 800 rpm without irrigation. 5. Osseointegration was successful in cases of 300, 800, 2,000 and 3,500 rpm with irrigation. But fibrous connective tissue formation was observed at the hollow of implant inserted with 5,000 rpm with irrigation.