Correction of Dose Distribution at Total Body Irradiation using Compensator
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Correction of Dose Distribution at Total Body Irradiation using Compensator
  • Correction of Dose Distribution at Total Body Irradiation using Compensator
저자명
김종식,조현상,김영곤,조정근,주상규,박영환,Kim. Jong Sik,Cho. Hyun Sang,Kim. Young Kon,Cho. Jung Keun,Ju. Sang Kyu,Park. Young Hwan
간행물명
大韓放射線治療技術學會誌
권/호정보
1997년|9권 1호|pp.87-93 (7 pages)
발행정보
대한방사선치료학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

The using of compensator is required to adjust the irregular dose distribution due to irregular thickness of the body in Total Body Irradiation. Aluminuim, copper or lead is generally used as compensator. In our study, we would like to introduce a result of the attenuation and compensation effect of radiation use compensator made by duralumin and its clinical use. The thickness of compensator was calculated by the attenustion of radiation, which was measured by polystyrene phantom and ionization chamber(farmer). The compensation effect of radiation was measured by diode detector. All of conditions were set as in real treatment, and the distanc from source to detector was 446 cm. We also made fixation of device to easily attach the compensator to LINAC. Beam spoiler was menufactured and placed on the patient to irradiate sufficient dose to the skin. diode detector were placed on head, neck, chest, umbilicus. pelvis and knee with each their entranced exit points, and datas of dose distribution were evaluated and compared in each points for eleven patients(Feb. 96-Feb. 97). The attenuation rate of irradiation by duralumin compensator was measured as $1.4\%$ in 2mm thickness. The mean attenuation rate was $1.3\%$ per 2mm as increasing the thickness gradually to 50 mm. By using duralunim compensator, dose distribution in each points of body was measured with ${pm}2.8\%$ by diode detectior. We could easily calculate the thickness of compensator by measuring the attenuation rate of radiation, remarkably reduce the irragularity of dose distribution duo to the thickness of body and magnify the effect of radiation therapy.