- 주파수의존형 전자근관장측정기의 개발과 정확도에 관한 연구 ; 전압차 자동보정형
- ㆍ 저자명
- 이승종,김덕원,남기창,Lee. Seung-Jong,Kim. Deok-Won,Nam. Ki-Chang
- ㆍ 간행물명
- 大韓齒科保存學會誌
- ㆍ 권/호정보
- 1997년|22권 2호|pp.809-819 (11 pages)
- ㆍ 발행정보
- 대한치과보존학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The purpose of this study was to develop an accurate frequency-dependent type apex locator. To compensate the impedence differences of different canal contents, voltage differences were pre-measured and stored for saline, HOsb1/$2O_2$ and NaOCl. The circuit was adjusted to recognize the different voltages of each solution, thereby the machine could automatically compensate the voltage differences. In the process of this study, the following informations were observed. 1. Most stable apical reading was obtained at 500Hz and 10kHz. 2. There were no differences in accuracy among three different frequency types, sign wave, triangular wave and rectangular wave(p>0.05). 3. Before the compensation of voltage differences, saline, H2O2 and NaOCl showed different readings at the apex even with the frequency-type (p<0.05). 4. After compensating the voltage difference, the accuracy was enhanced from + 0.42 to +0.32 in H2O2(p<0.05), from -0.34 to -0.12 in NaOCl(p<0.05).