- 흉부 팬텀 CR 영상의 화질평가 및 피폭선량 경감에 관한 연구
- ㆍ 저자명
- 김흥태,권수일,박일영,정환,Kim. Heung-Tae,Gwon. Su-Il,Park. Il-Yeong,Jeong. Hwan
- ㆍ 간행물명
- 대한디지털영상기술학회지
- ㆍ 권/호정보
- 1998년|4권 1호|pp.72-79 (8 pages)
- ㆍ 발행정보
- 대한디지털의료영상학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
With this experiment we have conclusion as follows. $cdot$ Total FCR-9501 are 8 units. $cdot$ All hospitals under study are conducting the high kvp radiographing with 115-120kvp and 4$sim$10mAs. $cdot$ The mean value of the chest exposure was 0.524$sim$0.301mGy. $cdot$ The percentage of the absorber finding from phantom was more the 85% with a range of 100$sim$136 kvp. $cdot$ The mean dose from phantom was 0.990mGy with 124 kvp and 16 mAs. $cdot$ There is no difference in the amount of information due to the change of kvp and mAs, but some coarseness of images can be seen in the magnified images. $cdot$ The resolving power was about 1.5/mm. $cdot$ The increase and decrease of 10$sim$30 kvp from common kvp has not influenced on the amount of information. $cdot$ The exposure could be reduced by 0.170$sim$0.570mGy.