- 유한요소법을 이용한 인공고관절 주대형태의 개선에 대한 연구
- ㆍ 저자명
- 윤경렬,원예연,이수훈
- ㆍ 간행물명
- 한국CAD/CAM학회논문집
- ㆍ 권/호정보
- 2000년|5권 2호|pp.122-126 (5 pages)
- ㆍ 발행정보
- 한국CAD/CAM학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
A major mechanical problem with total hip replacement is the loosening of the femoral component. The loss of proximal support, with firm fixation distally, has been thought to be a major caused of fatigue failure of femoral stems. While many causes have been proposed, the most frequently suggested cause of the calcar resorption is the disuse atrophy of the cortex of the calcar due to the stress shielding of the proximal bone by the metal femoral stem. In this research, the new-designed stem(modified collar stem) was considered which made a hole inside stem and had a 3 mm thickness. Using the 3-dimensional finite element methods, the common collar stem and the modified colla stem was modeled and analysed. Also, the two models was compared. The results showed that the modified collar stem decreased the stress-shielding and it made a effective load transfer at the entire femoral region.