- 골판톰을 이용한 골밀도측정 참조체의 결정
- ㆍ 저자명
- 김재덕,Kim. Jae-Duk
- ㆍ 간행물명
- 대한구강악안면방사선학회지
- ㆍ 권/호정보
- 2002년|32권 3호|pp.135-139 (5 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose: To determine the proper reference step wedge for digital Cu-Equivalent Image analyzing systems for measurement of bone density. Meterials and Methods : Radiograms of lumbar vertebrae phantom (1g/㎠) with 3 test copper step wedges of 0.03, 0.05 and, 0.1 mm thickness unit were taken and analyzed using NIH image software on a Macintosh personal computer. Measured densities of the lumbar areas in the Cu-Equivalent images made by utilizing 3 different copper stepwedges were compared with a known bone density. Results: The values of r2 for all copper equivalent images were over 0.99. The mean Cu-Eq value of lumbar in copper equivalent image made by a 0.1 mm copper stepwedge was 0.22 ± 0.06 mm and converted to hydroxyapatite density of 1.03 g/㎠. The stepwedges of 0.03 and 0.05 mm produced results having higher values than the actual known bone density. They did not show the blue and green color level that appeared in lumbar on color enhanced image. Conclusion : A copper stepwedge of adequate thickness and range of steps which can express the range of density of bone being measured should be used.