- 치과용 X-선 관구의 조정시간
- ㆍ 저자명
- 윤숙자,강병철,왕세명,고창성,Yoon. Suk-Ja,Kang. Byung-Cheol,Wang. Se-Myung,Koh. Chang-Sung
- ㆍ 간행물명
- 대한구강악안면방사선학회지
- ㆍ 권/호정보
- 2002년|32권 3호|pp.159-165 (7 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose: The aim of this study was to introduce a method of obtaining the oscillation graphs of the dental x-ray tube heads relative to time using an accelerometer. Materials and Methods: An Accelerometer, Piezotron type 8704B25 (Kistler Instrument Co., Amherst, NY, USA) was utilized to measure the horizontal oscillation of the x-ray tube head immediately after positioning the tube head for an intraoral radiograph. The signal from the sensor was transferred to a dynamic signal analyzer, which displayed the magnitude of the acceleration on the Y-axis and time lapse on the X -axis. The horizontal oscillation of the tube head was measured relative to time, and the settling time was also determined on the basis of the acceleration graphs for 6 wall type, 5 floor-fixed type, and 4 mobile type dental x-ray machines. Results : The oscillation graphs showed that tube head movement decreased rapidly over time. The settling time varied with x-ray machine types. Wall-type x-ray machines had a settling time of up to 6 seconds, 5 seconds for fixed floor-types, and 1 I seconds for the mobile-types. Conclusion: Using an accelerometer, we obtained the oscillation graphs of the dental x-ray tube head relative to time. The oscillation graph with time can guide the operator to decide upon the optimum exposure moment after x-ray tube head positioning for better radiographic resolution.