Development of Isotope Dilution-Liquid Chromatography/Tandem Mass Spectrometry as a Candidate Reference Method for the Determination of Acrylamide in Potato Chips
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Development of Isotope Dilution-Liquid Chromatography/Tandem Mass Spectrometry as a Candidate Reference Method for the Determination of Acrylamide in Potato Chips
  • Development of Isotope Dilution-Liquid Chromatography/Tandem Mass Spectrometry as a Candidate Reference Method for the Determination of Acrylamide in Potato Chips
저자명
Park. Sun-Young,Kim. Byung-Joo,So. Hun-Young,Kim. Yeong-Joon,Kim. Jeong-Kwon
간행물명
Bulletin of the Korean Chemical Society
권/호정보
2007년|28권 5호|pp.737-744 (8 pages)
발행정보
대한화학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

An isotope dilution-liquid chromatography/tandem mass spectrometric method was developed as a candidate reference method for the accurate determination of acrylamide in potato chips, starch-rich foodstuff cooked at high temperature. Sample was spiked with 13C3-acrylamide and then extracted with water. The extract was further cleaned up with an Oasis HLB solid-phase extraction (SPE) cartridge and an Oasis mixed-phase cation exchange (MCX) SPE cartridge. The extract was analyzed by using LC/ESI/Tandem MS in positive ion mode. LC with a medium reversed-phase (C4) column was optimized to obtain adequate chromatographic retention and separation of acrylamide. MS was operated to selectively monitor [M+H]+ ions of the analyte and its isotope analogue at m/z 72 and m/z 75, respectively. Sample was also analyzed by the LC/MS with selectively monitoring the collisionally induced dissociation channels of m/z 72 → m/z 55 and m/z 75 → 58. Compared to the LC/MS chromatograms, the LC/MS/MS chromatograms showed substantially reduced background chemical noises coming from solvent clusters formed during ESI spray processes and interferences from sample matrix. Repeatability and reproducibility studies showed that the LC/MS/MS method is a reliable and reproducible method which can provide a typical method precision of 1.0% while the LC/MS results are influenced by chemical interferences.