- 생명신호 측정용 반사형 광용적맥파 측정기의 움직임에 의한 신호왜곡 제거
- ㆍ 저자명
- 한효녕,이연주,김정식,김정,Han. Hyo-Nyoung,Lee. Yun-Joo,Kim. Jung-Sik,Kim. Jung
- ㆍ 간행물명
- 한국정밀공학회지
- ㆍ 권/호정보
- 2009년|26권 12호|pp.146-153 (8 pages)
- ㆍ 발행정보
- 한국정밀공학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
One of the most important issues in the wearable healthcare sensors is to minimize the motion artifacts in the vital signals for continuous monitoring. This paper presents a reflected type photoplethysmograph (PPG) sensor for monitoring heart rates at the artery of the wrist. Active noise cancellation algorithm was applied to compensate the distorted signals by motions with Least Mean Square (LMS) adaptive filter algorithms, using acceleration signals from a MEMS accelerometer. Experiments with a watch type PPG sensor were performed to validate the proposed algorithm during typical daily motions such as walking and running. The developed sensor is suitable for ubiquitous healthcare system and monitoring vital arterial signals during surgery.