- 이동형 X-ray 발생장치를 이용한 복부 촬영 시 공간 선량률에 관한 연구
- ㆍ 저자명
- 박창희,Park. Chang-Hee
- ㆍ 간행물명
- Korean journal of digital imaging in medicine : KJDIM
- ㆍ 권/호정보
- 2010년|12권 2호|pp.97-101 (5 pages)
- ㆍ 발행정보
- 대한디지털의료영상학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
This experimental study is carried out one of the General Hospital in Kyungbok providence. Abdomen Phantom being located Anterior-posterior(AP) position on portable bed, and the portable X-ray generating device was placed the phantom at $-90^{circ}$ direction. The experiment were set 65 kVp, 10 mAs, $10{ imes}10;cm^2$, 100 cm(FOD) for the measurement. Digital proportional counting tube survey meter was used for measuring the space scatter dose. Measurement points of horizontal distribution was set up at $30^{circ}$ interval by increasing 50 cm radius of upside, downside, left and right. Vertical distribution of measurement points were set up for the vertical plane with a radius of at $30^{circ}$ intervals with 50cm increments. It is concluded that longer distance from the soure of X-ray significantly decrease radiation dose to the patient and use of the radiation protection device should be applied in clinical practice to reduce dose to the patient.