  • QUANTITATIVE DATA TO SHOW EFFECTS OF GEOMETRIC ERRORS AND DOSE GRADIENTS ON DOSE DIFFERENCE FOR IMRT DOSE QUALITY ASSURANCE MEASUREMENTS
  • QUANTITATIVE DATA TO SHOW EFFECTS OF GEOMETRIC ERRORS AND DOSE GRADIENTS ON DOSE DIFFERENCE FOR IMRT DOSE QUALITY ASSURANCE MEASUREMENTS
저자명
Park. So-Yeon,Park. Jong-Min,Ye. Sung-Joon
간행물명
방사선방어학회지
권/호정보
2011년|36권 4호|pp.183-189 (7 pages)
발행정보
대한방사선방어학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

To quantitatively evaluate how setup errors in conjunction with dose gradients contribute to the error in IMRT dose quality assurance (DQA) measurements. The control group consisted of 5 DQA plans of which all individual field dose differences were less than ${pm}5%$. On the contrary, the examination group was composed of 16 DQA plans where any individual field dose difference was larger than ${pm}10%$ even though their total dose differences were less than ${pm}5%$. The difference in 3D dose gradients between the two groups was estimated in a cube of $6{ imes}6{ imes}6;mm^3$ centered at the verification point. Under the assumption that setup errors existed during the DQA measurements of the examination group, a three dimensional offset point inside the cube was sought out, where the individual field dose difference was minimized. The average dose gradients of the control group along the x, y, and z axes were 0.21, 0.20, and 0.15 $cGy{cdot}mm^{-1}$, respectively, while those of the examination group were 0.64, 0.48, and 0.28 $cGy{cdot}mm^{-1}$, respectively. All 16 plans of the examination group had their own 3D offset points in the cube. The individual field dose differences recalculated at the offset points were mostly diminished and thus the average values of total and individual field dose differences were reduced from 3.1% to 2.2% and 15.4% to 2.2%, respectively. The offset distribution turned out to be random in the 3D coordinate. This study provided the quantitative data that support the large individual field dose difference mainly stems from possible geometric errors (e.g., random setup errors) under the influence of steep dose gradients of IMRT field.