  • Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test
  • Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test
저자명
Bayarchimeg. Dorjpalam,Namgoong. Hee,Kim. Byung Kook,Kim. Myung Duk,Kim. Sungtae,Kim. Tae-Il,Seol. Yang Jo,Lee. Yong Moo,Ku. You
간행물명
Journal of periodontal & implant science
권/호정보
2013년|43권 1호|pp.30-36 (7 pages)
발행정보
대한치주과학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Purpose: Implant stability at the time of surgery is crucial for the long-term success of dental implants. Primary stability is considered of paramount importance to achieve osseointegration. The purpose of the present study was to investigate the correlation between the insertion torque and primary stability of dental implants using artificial bone blocks with different bone densities and compositions to mimic different circumstances that are encountered in routine daily clinical settings. Methods: In order to validate the objectives, various sized holes were made in bone blocks with different bone densities (#10, #20, #30, #40, and #50) using a surgical drill and insertion torque together with implant stability quotient (ISQ) values that were measured using the Osstell Mentor. The experimental groups under evaluation were subdivided into 5 subgroups according to the circumstances. Results: In group 1, the mean insertion torque and ISQ values increased as the density of the bone blocks increased. For group 2, the mean insertion torque values decreased as the final drill size expanded, but this was not the case for the ISQ values. The mean insertion torque values in group 3 increased with the thickness of the cortical bone, and the same was true for the ISQ values. For group 4, the mean insertion torque values increased as the cancellous bone density increased, but the correlation with the ISQ values was weak. Finally, in group 5, the mean insertion torque decreased as the final drill size increased, but the correlation with the ISQ value was weak. Conclusions: Within the limitations of the study, it was concluded that primary stability does not simply depend on the insertion torque, but also on the bone quality.