Measurement Uncertainty of Nicotine in Environmental Tobacco Smoke (ETS)
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Measurement Uncertainty of Nicotine in Environmental Tobacco Smoke (ETS)
저자명
Lee. Jeong-Il,Lee. Cheol Min,Shim. In-Keun,Kim. Seong-Mi,Lee. Woo-Seok,Kim. Yoon-Shin
간행물명
Bulletin of the Korean Chemical Society
권/호정보
2013년|34권 8호|pp.2394-2398 (5 pages)
발행정보
대한화학회
파일정보
정기간행물|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Nicotine is the main component of environmental tobacco smoke, and its presence in indoor air is widely used as a secondhand-smoke indicator. Environmental tobacco smoke is a major source of indoor air pollution, but sufficient investigation of the uncertainty of its measurement, which mirrors the reliability of nicotine measurement, has not been performed. We calculated the uncertainty of measurement of indoor air nicotine concentration at low, medium, and high concentrations of 11.3798, 10.1977, $98.3768{mu}g/m^3$, respectively, and we employed the Guide to the Expression of Uncertainty in Measurements (GUM), proposed by the International Organization for Standardization (ISO). The factors considered in determining the uncertainty were uncertainty of the calibration curve (calibration curve and repeated measurements), desorption efficiency, extraction volume, and sampling airflow (accuracy and acceptable limits of flowmeter). The measurement uncertainty was highest at low concentrations; the expanded measurement uncertainty is $0.9435{mu}g/m^3$ and is represented as a relative uncertainty of 63.38%. At medium and high (concentrations, the relative uncertainty was 13.1% and 9.1%, respectively. The uncertainty of the calibration curve was largest for low indoor nicotine concentrations. To increase reliability of measurement in assessing the effect of secondhand smoke, measures such as increasing the sample injection rate ($1{mu}L$ or more), increasing sampling volume to increase collected nicotine, and using gas chromatography-mass spectrometry (GC/MS) or GC/MS/MS, which has a lower quantitation threshold, rather than gas chromatography with nitrogen phosphorous detector, should be considered.