- CTD 교정 시스템 구축 및 불확도 평가
- ㆍ 저자명
- 이정한,황근춘,김은수,이승훈,Lee. Jung-Han,Hwang. Keun-Choon,Kim. Eun-Soo,Lee. Seung-Hun
- ㆍ 간행물명
- Ocean and polar research
- ㆍ 권/호정보
- 2014년|36권 1호|pp.77-85 (9 pages)
- ㆍ 발행정보
- 한국해양과학기술원
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The quality control of ocean observations data is becoming a major issue as real-time observational data and information services have increased recently. Therefore, it is necessary for oceanographic instruments to calibrate. In this paper, we first introduce the CTD calibration system and traceability. Next, CTD calibration procedures and estimation of uncertainty of measurement are described. The expanded uncertainty (k = 2) of the temperature, pressure and conductivity are 0.$0.003^{circ}C$, $6.0{ imes}10^{-5}$ and 0.006 mS/cm respectively. Finally, the excellence of CTD calibration and its measurement capability has been proven by comparing the inter-calibration result of KIOST and Sea-Bird Electronics (SBE). CTD calibration residuals are less than ${pm}0.0001^{circ}C$, ${pm}0.001$ MPa, ${pm}0.0001$ S/m for SBE 3plus temperature sensor, SBE 19plus pressure sensor and SBE 4C conductivity sensor respectively.