- 슬관절 재전치환술용 경골 삽입물 형상이 주대 말단부의 접촉압력과 응력 분포에 미치는 영향
- ㆍ 저자명
- 김윤혁,권오수,박시몬,Kim. Yoon-Hyuk,Kwon. Oh-Soo,Park. Si-Mon
- ㆍ 간행물명
- 한국정밀공학회지
- ㆍ 권/호정보
- 2006년|23권 11호|pp.126-134 (9 pages)
- ㆍ 발행정보
- 한국정밀공학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
In this study, the effect of stem-end design on contact pressure and stress distribution in revision TKR was investigated using finite element method. The finite element model of tibia, including the cortical bone, the cancellous bone and canal, was developed based on CT images. The implant models with various stem lengths, diameters, friction coefficients, and press-fit effects were considered. The results showed that the longer stem length, the stronger press-fit, the bigger stem diameter, and the higher friction coefficient increased both peak contact pressure and Von-Mises stress distributions. The results supported the clinical hypothesis that peak contact pressure and stress are related to the stem end pain. The results of this study will be useful to design the stem and reduce the end-of-stem pain in revision TKR.